Compact Inexpensive High Resolution Retarding Field Energy Analyzer
- 1 March 1972
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 43 (3) , 427-430
- https://doi.org/10.1063/1.1685653
Abstract
A compact energy analyzer which has the four‐grid construction as is widely used in LEED—Auger systems was developed to analyze secondary electrons. It is about 4.0 cm diam ×3.5 cm in length, with an entrance aperture of 5.6 mm diam and an angular aperture of about 10°. To get a uniform retarding field, a pair of fringing rings were added at the edge of the first and the second grids. The first and the second grids (gold coated tungsten, 100 mesh) were shaped into a portion of a sphere. The distance between the source of secondary electrons and the first or the second grid was made somewhat longer than their radius of curvature. The energy resolution was studied by adjusting the distance between the grids, and a value of ΔE/E of about 0.09% was attained.Keywords
This publication has 7 references indexed in Scilit:
- Magnetic Field Uniformity around Near-Helmholtz Coil ConfigurationsReview of Scientific Instruments, 1969
- HIGH SENSITIVITY AUGER ELECTRON SPECTROMETERApplied Physics Letters, 1969
- Instrumental Effects of the Retarding Grids in a LEED ApparatusReview of Scientific Instruments, 1969
- Resolution and Sensitivity Considerations of an Auger Electron Spectrometer Based on Display LEED OpticsReview of Scientific Instruments, 1969
- Fast Response PicoammeterReview of Scientific Instruments, 1966
- Double Grid Repeller System to Improve Electron Resolution in Low Energy Electron Diffraction EquipmentReview of Scientific Instruments, 1965