Space charge effects on relative peak heights in fourier transform-ion cyclotron resonance spectra
- 1 October 1992
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 3 (7) , 734-741
- https://doi.org/10.1016/1044-0305(92)87086-e
Abstract
Ion trajectory calculations have confirmed that space charge interactions can be a source for mass discrimination seen in Fourier transform-ion cyclotron resonance (FT-ICR) spectra. As compared with the previously recognized mechanism of z-axis excitation, ion-ion repulsion is a mechanism which specifically affects relative peak heights of ions close in mass, and is most severe for low excitation radiofrequency (rf) amplitudes. In this mechanism, Coulomb repulsion significantly perturbs the motion of the ion clouds during excitation and alters the final cyclotron orbital radii. Under these conditions peak heights do not accurately reflect the true ion abundances in the FT-ICR spectrometer. Mass discrimination can be minimized by using low numbers of ions, low ion densities, and a short, high amplitude rf excitation waveform. Experimental observation of the relative peak heights of the m/z 91, 92, and 134 ions in n-butylbenzene gives quantitative confirmation of the results of the trajectory calculations. Chirp, SWIFT, and impulse excitation were modeled: impulse excitation was found to be most effective in minimizing the effects of space charge interactions.Keywords
This publication has 17 references indexed in Scilit:
- Elimination of z-ejection in Fourier-transform ion cyclotron resonance mass spectrometry by radio/frequency electric field shimmingAnalytical Chemistry, 1990
- Toward quantitation of ion/molecule kinetics in fourier transform ion cyclotron resonance mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Excitation of the z-motion of ions in a cubic icr cellInternational Journal of Mass Spectrometry and Ion Processes, 1988
- Coupling of axial and radial motions in ICR cells during excitationInternational Journal of Mass Spectrometry and Ion Processes, 1986
- Mass-dependent z-excitation of ions in cubic traps used in FTMSInternational Journal of Mass Spectrometry and Ion Processes, 1986
- Relation of signal sensitivity and ion z-motion in cubic cells. Theory and implication for ion kinetic studiesInternational Journal of Mass Spectrometry and Ion Processes, 1986
- Coulomb broadening in Fourier transform ion cyclotron resonance mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1986
- The effect of ion position on ICR signal strengthInternational Journal of Mass Spectrometry and Ion Processes, 1984
- Theory of Fourier transform ion cyclotron resonance mass spectroscopy: Response to frequency-sweep excitationThe Journal of Chemical Physics, 1980
- Fourier transform ion cyclotron resonance spectroscopyChemical Physics Letters, 1974