A Randomized, Controlled Trial Comparing Two Different Continuous Positive Airway Pressure Systems for the Successful Extubation of Extremely Low Birth Weight Infants
- 1 November 2003
- journal article
- clinical trial
- Published by American Academy of Pediatrics (AAP) in Pediatrics
- Vol. 112 (5) , 1031-1038
- https://doi.org/10.1542/peds.112.5.1031
Abstract
Objective. To determine whether the use of the Infant Flow continuous positive airway pressure (IF CPAP) system reduces the rate of extubation failure among extremely low birth weight (ELBW) infants (infants with birth weight Methods. A prospective, unmasked, randomized, controlled clinical trial was conducted in 162 eligible intubated ELBW infants who were hospitalized in 2 intensive care nurseries in Winston-Salem, North Carolina, between July 1997 and November 2000. Successful extubation was defined as no need for reintubation for any reason for at least 7 days after the first extubation attempt. Results. The individual extubation success rates were 61.9% (52 of 84) in the conventional CPAP group and 61.5% (48 of 78) in the IF CPAP group. There were no significant differences in the extubation success rate in any birth weight subset between the 2 cohorts. The most common cause of extubation failure was apnea/bradycardia. Infants who were randomized to IF CPAP had fewer days on supplemental O2 and shorter hospital stays. Conclusions. Extubation failure is a common problem, occurring in nearly 40% of ELBW infants who require mechanical ventilation. IF CPAP was as effective but no more effective than conventional CPAP in preventing extubation failure among ELBW infants. New strategies are needed to identify predictors of extubation success and to treat apnea/bradycardia, the most common cause of extubation failure, thereby reducing the likelihood of prolonged intubation in this high-risk cohort of premature infants.Keywords
This publication has 31 references indexed in Scilit:
- Continuous distending pressure for respiratory distress in preterm infantsPublished by Wiley ,2002
- Extubation from low-rate intermittent positive airway pressure versus extubation after a trial of endotracheal continuous positive airway pressure in intubated preterm infantsCochrane Database of Systematic Reviews, 2001
- Nasal intermittent positive pressure ventilation (NIPPV) versus nasal continuous positive airway pressure (NCPAP) for preterm neonates after extubationPublished by Wiley ,2001
- Methylxanthine therapy in premature infants: Sound practice, disaster, or fruitless byway?The Journal of Pediatrics, 1999
- Application of nasal continuous positive airway pressure to early extubation in very low birthweight infants.Archives of Disease in Childhood: Fetal & Neonatal, 1995
- Does continuous positive airway pressure (CPAP) during weaning from intermittent mandatory ventilation in very low birth weight infants have risks or benefits? A controlled trialPediatric Pulmonology, 1995
- Randomized, controlled trial of nasopharyngeal continuous positive airway pressure in the extubation of very low birth weight infantsThe Journal of Pediatrics, 1994
- Nasal continuous positive airway pressure: experiences with a new technical approachActa Paediatrica, 1993
- A new device for administration of nasal continuous positive airway pressure in the newbornCritical Care Medicine, 1988
- Effects on respiration of CPAP immediately after extubation in the very preterm infantPediatric Pulmonology, 1988