Preventing Medical Errors in Point-of-Care Testing
Open Access
- 1 October 2001
- journal article
- Published by Archives of Pathology and Laboratory Medicine in Archives of Pathology & Laboratory Medicine
- Vol. 125 (10) , 1307-1315
- https://doi.org/10.5858/2001-125-1307-pmeipo
Abstract
Objective.—To prevent medical errors, improve user performance, and enhance the quality, safety, and connectivity (bidirectional communication) of point-of-care testing. Participants.—Group A included 37 multidisciplinary experts in point-of-care testing programs in critical care and other hospital disciplines. Group B included 175 professional point-of-care managers, specialists, clinicians, and researchers. The total number of participants equaled 212. Evidence.—This study followed a systems approach. Expert specifications for prevention of medical errors were incorporated into the designs of security, validation, performance, and emergency systems. Additional safeguards need to be implemented through instrument software options and point-of-care coordinators. Connectivity will be facilitated by standards that eliminate deficiencies in instrument communication and device compatibility. Assessment of control features on handheld, portable, and transportable point-of-care instruments shows that current error reduction features lag behind needs. Consensus Process.—Step 1: United States national survey and collation of group A expert requirements for security, validation, and performance. Step 2: Design of parallel systems for these functions. Step 3: Written critique and improvement of the error-prevention systems during 4 successive presentations to group B participants over 9 months until system designs stabilized into final consensus form. Conclusions.—The consensus process produced 6 conclusions for preventing medical errors in point-of-care testing: (1) adopt operator certification and validation in point-of-care testing programs; (2) implement security, validation, performance, and emergency systems on existing and new devices; (3) require flexible, user-defined error-prevention system options on instruments as a prerequisite to federal licensing of new diagnostic tests and devices; (4) integrate connectivity standards for bidirectional information exchange; (5) preserve fast therapeutic turnaround time of point-of-care test results; and (6) monitor invalid use, operator competence, quality compliance, and other performance improvement indices to reduce errors, thereby focusing on patient outcomes. (Arch Pathol Lab Med. 2001;1307–1315)Keywords
This publication has 19 references indexed in Scilit:
- Point‐of‐Care TestingPublished by Wiley ,2000
- Deaths Due to Medical Errors Are Exaggerated in Institute of Medicine ReportJAMA, 2000
- Institute of Medicine Medical Error Figures Are Not ExaggeratedJAMA, 2000
- Point-of-Care Testing: Millennium Technology for Critical CareLaboratory Medicine, 2000
- Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose AnalyzerAmerican Journal of Clinical Pathology, 2000
- Frequency of Unacceptable Results in Point-of-Care TestingArchives of Pathology & Laboratory Medicine, 1999
- Continuous Quality Improvement for Point-of-Care Testing Using Background Monitoring of Duplicate SpecimensArchives of Pathology & Laboratory Medicine, 1999
- The Laboratory-Clinical InterfaceChest, 1999
- Error in MedicineJAMA, 1994
- Critical Limits for Urgent Clinician Notification at US Medical CentersPublished by American Medical Association (AMA) ,1990