TLD, diode and Monte Carlo dosimetry of an192Ir source for high dose-rate brachytherapy
- 1 December 1995
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 40 (12) , 2015-2036
- https://doi.org/10.1088/0031-9155/40/12/002
Abstract
Very few dosimetry data are available for the current generation of high-dose-rate (HDR) 192Ir sources, which have broad application in remotely afterloaded brachytherapy. We have measured the two-dimensional dose rate distribution around a microSelectron-HDR source and used the results to validate Monte Carlo simulations. Thermoluminescent dosimeters (TLDs) in solid-water phantoms were used to measure the transverse-axis dose rates in the distance range 0.5-10 cm and the polar dose-rate profiles at 1.5, 3 and 5 cm distance from the source. At close distances, 2-40 mm from the HDR source, we performed transverse axis dose-rate measurements with a Si diode in water. We performed diode measurements at the same distances also for a pulsed dose-rate (PDR) source to compare the results for 192Ir sources with different encapsulation. Both the HDR and the PDR sources were decayed, separated from their cables and calibrated prior to the measurements. The measured dose rates were compared with Monte Carlo photon transport calculations, which realistically modelled the experimental and source geometry at each measurement point. Agreement between Monte Carlo photon transport absolute dose-rate calculations and measurements was, on average, within 5%. From the transverse-axis experimental data, we deduced a value for the dose-rate constant lambda 0 of 192Ir HDR sources of 1.14 cGy h-1 U-1 +/- 5%. This value agrees within the experimental error with the Monte Carlo estimate of 1.115 cGy h-1 U-1 +/- 0.5%. Excellent agreement with previously measured anisotropy functions was observed. Higher anisotropy is observed for the point at 0 degree along the source cable for which no previous data have been reported.Keywords
This publication has 23 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsPublished by Elsevier ,2002
- Quantitative verification of 192Ir PDR and HDR source structure by pin‐hole autoradiographyMedical Physics, 1995
- Instrumentation and dosimeter‐size artifacts in quantitative thermoluminescence dosimetry of low‐dose fieldsMedical Physics, 1995
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Autoradiography for iodine-125 seedsInternational Journal of Radiation Oncology*Biology*Physics, 1993
- Volume-based geometric modeling for radiation transport calculationsMedical Physics, 1992
- Calibration of 192Ir high‐dose‐rate afterloading systemsMedical Physics, 1991
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983
- Specific γ-ray constant and exposure rate constant of 192 IrMedical Physics, 1979
- The Effective Attenuation in Water of the Gamma Rays of Gold 198, Iridium 192, Cesium 137, Radium 226, and Cobalt 60Radiology, 1968