The Effect of Source Size upon Integral Dose in Teletherapy Units
- 1 July 1954
- journal article
- research article
- Published by Taylor & Francis in Acta Radiologica
- Vol. 42 (1) , 65-74
- https://doi.org/10.3109/00016925409175097
Abstract
A simple formula is deduced on theoretical grounds for calculating the effect of source size upon integral dose in teletherapy units. Measurements of integral dose using the ferrous sulfate system of chemical dosimetry were made and confirm the theoretical predictions. Particular attention is drawn to the critical influence upon integral dose of the distance from the skin to the defining aperture. If this distance is not too large, calculations show that the source-size in large teletherapy units need not increase integral doses above the levels that have been tolerated in the past.Keywords
This publication has 1 reference indexed in Scilit:
- The Installation of Co60 in a Gamma Ray Beam UnitActa Radiologica, 1952