Characteristics of the high‐energy photon beam of a 25‐MeV accelerator
- 1 January 1988
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 15 (1) , 82-87
- https://doi.org/10.1118/1.596163
Abstract
The CGR Saturne 25 is an isocentrically mounted standing wave medical linear accelerator that produces dual-energy photon beams and a scanned electron beam with six selectable energies between 4 and 25 MeV. The highest energy photon beam is nominally referred to as 23 MV. For this beam the mean energy of the accelerated electron beam on the 1.3 radiation length (4 mm) tungsten X-ray target is found to be approximately 21 MeV, with the energy acceptance stated to be .+-. 5%. The electron beam traverses a 270.degree. bending magnet upstream of the X-ray production target. The resulting bremsstrahlung beam passes through a combination steel and lead flattening filter, 4-cm maximum thickness. Dosimetric data for the 23-MV beam are presented with respect to rectanglular field output factor, depth of maximum dose as a function of field size, surface and buildup dose, central axis percent depth dose, tissue-phantom ratios, beam profile, applicability of inverse square, and block transmission. Some data are also presented on the effect of different flattening filter designs on apparent beam energy.Keywords
This publication has 3 references indexed in Scilit:
- Output factor constituents of a high‐energy photon beamMedical Physics, 1988
- Head scatter data for several linear accelerators (4-18 MV)Medical Physics, 1986
- Characteristics of an 18 MV photon beam from a Therac 20 Medical Linear AcceleratorMedical Physics, 1981