Characteristics and quality assurance of a dedicated open 0.23 T MRI for radiation therapy simulation
- 23 October 2002
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 29 (11) , 2541-2547
- https://doi.org/10.1118/1.1513991
Abstract
A commercially available open MRI unit is under routine use for radiation therapy simulation. The effects of a gradient distortion correction (GDC) program used to post process the images were assessed by comparison with the known geometry of a phantom. The GDC reduced the magnitude of the distortions at the periphery of the axial images from 12 mm to 2 mm horizontally along the central axis and distortions exceeding 20 mm were reduced to as little as 2 mm at the image periphery. Coronal and sagittal scans produced similar results. Coalescing these data into distortion as a function of radial distance, we found that for radial distances of <10 cm, the distortion after GDC was <2 mm and for radial distances up to 20 cm, the distortion was <5 mm. The dosimetric errors resulting from homogeneous dose calculations with this level of distortion of the external contour is <2%. A set of triangulation lasers has been added to establish a virtual isocenter for convenient setup and marking of patients and phantoms. Repeated measurements of geometric phantoms over several months showed variations in position between the virtual isocenter and the magnetic isocenter were constrained to <2 mm. Additionally, the interscan variations of 12 randomly selected points in space defined by a rectangular grid phantom was found to be within the intraobserver error of approximately 1 mm in the coronal, sagittal, and transverse planes. Thus, the open MRI has sufficient geometric accuracy for most radiation therapy planning and is temporally stable.Keywords
This publication has 22 references indexed in Scilit:
- Routine use of MRI and CT simulation for treatment planning of intensity modulated radiation therapy (IMRT) in prostate cancerInternational Journal of Radiation Oncology*Biology*Physics, 2001
- Einsatzmöglichkeiten eines offenen magnetresonanztomographen in der therapiesimulation und dreidimensionalen bestrahlungsplanungStrahlentherapie und Onkologie, 1999
- Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostateInternational Journal of Radiation Oncology*Biology*Physics, 1997
- Technical Note. Effect of Geometrical Distortion Correction in MR on Image Registration AccuracyJournal of Computer Assisted Tomography, 1996
- An evaluation of setup uncertainties for patients treated to pelvic sitesInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Gradient coil design: A review of methodsMagnetic Resonance Imaging, 1993
- Analysis of machine-dependent and object-induced geometric distortion in 2DFT MR imagingMagnetic Resonance Imaging, 1992
- A quantitative assessment of the addition of MRI to CT-based, 3-D treatment planning of brain tumorsRadiotherapy and Oncology, 1992
- Sagittal and coronal planes from MRI for treatment planning in tumors of brain, head and neck: MRI assisted simulationRadiotherapy and Oncology, 1992
- Integration of magnetic resonance imaging into radiation therapy treatment planning: i. technical considerationsInternational Journal of Radiation Oncology*Biology*Physics, 1987