Hydride Generation Interface for Speciation Analysis Coupling Capillary Electrophoresis to Inductively Coupled Plasma Mass Spectrometry

Abstract
A novel hydride generation (HG) interface for coupling capillary electrophoresis (CE) with inductively coupled plasma mass spectrometry (ICPMS) is presented in this work. The CE-HG-ICPMS interface was applied to the separation and quantitation of common arsenic species. Lack of a commercially available HG interface for CE-ICPMS led to a three concentric tube design allowing alleviation of back pressure commonly observed in CE-HG-ICPMS. Due to the high sensitivity and element-specific detection of ICPMS, quantitative analysis of As(III), As(V), monomethylarsonic acid, and dimethylarsinic acid was achieved. Optimization of CE separation conditions resulted in the use of 20 mmol L-1 sodium borate with 2% osmotic flow modifier (pH 9.0) and −20 kV applied potential for baseline resolution of each arsenic species in the shortest time. Hydride generation conditions were optimized through multiple electrophoretic separation analyses with 5% HCl and 3% NaBH4 (in 0.2% NaOH) determined to be the optimum conditions. After completion of system optimization, detection limits obtained for the arsenic species were less than 40 ng L-1 with electromigration time precision less than 1% within a total analysis time of 9.0 min. Finally, the interface was used for speciation analysis of arsenic in river and tap water samples.

This publication has 17 references indexed in Scilit: