Pulsed Nuclear Resonance Apparatus
- 1 March 1964
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 35 (3) , 316-333
- https://doi.org/10.1063/1.1718815
Abstract
The description of a complete pulsed nuclear resonance spectrometer suitable for experiments in solids is given. Its major features are the ability to operate at any frequency between 2 and 30 Mc, an rf pulse output power of 5 kW over most of this range, a receiving system recovery time of 5–10 μsec at 10 Mc, a sensitivity usually limited by thermal noise in the receiver coil, use of phase coherent detection, the availability of a wide variety of pulse sequences, and good long and short term stability of the entire system. Brief instructions for operating the apparatus are included. A method of using the ``boxcar'' integrator to record directly the steady‐state absorption (χ″) or dispersion (χ′) is explained.Keywords
This publication has 9 references indexed in Scilit:
- ``Boxcar'' Integrator with Long Holding TimesReview of Scientific Instruments, 1961
- rf Gate with 109 Carrier SuppressionReview of Scientific Instruments, 1961
- Nuclear Quadrupole Spin-Lattice Relaxation in Alkali HalidesPhysical Review B, 1960
- Multiple Echoes in SolidsPhysical Review B, 1958
- Free-Induction Decays in SolidsPhysical Review B, 1957
- Nuclear Spin Relaxation in Alkali MetalsPhysical Review B, 1955
- Effects of Diffusion on Free Precession in Nuclear Magnetic Resonance ExperimentsPhysical Review B, 1954
- Spin EchoesPhysical Review B, 1950
- Relaxation Effects in Nuclear Magnetic Resonance AbsorptionPhysical Review B, 1948