Malaria detection with the Sysmex XE-2100 hematology analyzer using pseudoeosinophilia and abnormal WBC scattergram
- 22 April 2008
- journal article
- research article
- Published by Springer Nature in Annals of Hematology
- Vol. 87 (9) , 755-759
- https://doi.org/10.1007/s00277-008-0486-8
Abstract
Recent investigation using the Sysmex XE-2100 hematology analyzer (Sysmex Corporation, Japan) has demonstrated erroneously high eosinophil counts and abnormal white blood cell (WBC) scattergrams in malaria cases. This study was conducted to assess the diagnostic efficiency of the Sysmex XE-2100 analyzer for malaria. One hundred forty-four patients initially diagnosed with Plasmodium vivax infection, 319 patients with febrile illness, and 24 patients who underwent malaria treatment were analyzed. Atypical features on Sysmex XE-2100 analyzer were categorized as pseudoeosinophilia (a gap of more than 5% in eosinophil counts between the Sysmex XE-2100 analyzer and microscopic examination) and abnormal WBC scattergram. Pseudoeosinophilia or abnormal WBC scattergram were detected in 100 of 144 malaria-positive samples (sensitivity 69.4%, specificity 100%). The samples with pseudoeosinophilia or abnormal WBC scattergrams showed significantly higher parasite counts than the samples without pseudoeosinophilia or an abnormal WBC scattergram (P < 0.05). All 24 samples from patients for whom the malaria smear was repeated after malaria treatment was initiated showed a normalized eosinophil count and a normal WBC histogram. In conclusion, attention to differential count and WBC scattergrams provided by the Sysmex XE-2100 would be a valuable tool in malaria detection.Keywords
This publication has 19 references indexed in Scilit:
- Haemozoin: from melatonin pigment to drug target, diagnostic tool, and immune modulatorThe Lancet Infectious Diseases, 2007
- Three Cases of Pseudoeosinophilia Associated with Malaria Determined in the Sysmex XE-2100 Automated Hematology AnalyzerAnnals of Laboratory Medicine, 2006
- Sensitivity of laser light depolarization analysis for detection of malaria in blood samplesJournal of Medical Microbiology, 2005
- Pseudoeosinophilia associated with malaria infection determined in the Sysmex XE-2100 hematology analyzerAnnals of Hematology, 2004
- Sensitivity of hemozoin detection by automated flow cytometry in non‐ and semi‐immune malaria patientsCytometry Part B: Clinical Cytometry, 2003
- Avoiding Misdiagnosis of Malaria: A Novel Automated Method Allows Specific Diagnosis, even in the Absence of Clinical SuspicionEmerging Infectious Diseases, 1999
- Re-emerging Plasmodium vivax malaria in the Republic of KoreaThe Korean Journal of Parasitology, 1999
- Malaria diagnosis: standardization of a polymerase chain reaction for the detection of Plasmodium falciparum parasites in individuals with low-grade parasitemiaZeitschrift Fur Parasitenkunde-Parasitology Research, 1996
- Malaria pigment in leucocytesTransactions of the Royal Society of Tropical Medicine and Hygiene, 1995
- Intraleucocytic malaria pigment and prognosis in severe malariaTransactions of the Royal Society of Tropical Medicine and Hygiene, 1995