Measurement and correction of transmitter and receiver induced nonuniformities in vivo
- 27 January 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 53 (2) , 408-417
- https://doi.org/10.1002/mrm.20354
Abstract
Signal intensity nonuniformities in high field MR imaging limit the ability of MRI to provide quantitative information and can negatively impact diagnostic scan quality. In this paper, a simple method is described for correcting these effects based on in vivo measurement of the transmission field B1+ and reception sensitivity maps. These maps can be obtained in vivo with either gradient echo (GE) or spin echo (SE) imaging sequences, but the SE approach exhibits an advantage over the GE approach for correcting images over a range of flip angles. In a uniform phantom, this approach reduced the ratio of the signal SD to its mean from around 30% before correction to approximately 6% for the SE approach and 9% for the GE approach after correction. The application of the SE approach for correcting intensity nonuniformities is demonstrated in vivo with human brain images obtained using a conventional spin echo sequence at 3.0 T. Furthermore, it is also shown that this in vivo B1+ and reception sensitivity mapping can be performed using segmented echo planar imaging sequences providing acquisition times of less than 2 min. Although the correction presented here is demonstrated with a simultaneous transmit and receive volume coil, it can be extended to the case of separate transmission and reception coils, including surface and phase array coils. Magn Reson Med 53:408–417, 2005.Keywords
This publication has 27 references indexed in Scilit:
- Polarization of the RF field in a human head at high field: A study with a quadrature surface coil at 7.0 TMagnetic Resonance in Medicine, 2002
- Analysis of wave behavior in lossy dielectric samples at high fieldMagnetic Resonance in Medicine, 2002
- Qualitative and Quantitative Evaluation of Six Algorithms for Correcting Intensity Nonuniformity EffectsNeuroImage, 2001
- The principle of reciprocity in signal strength calculations?A mathematical guideConcepts in Magnetic Resonance, 2000
- A simple method for investigating the effects of non-uniformity of radiofrequency transmission and radiofrequency reception in MRI.The British Journal of Radiology, 1998
- MRI segmentation: Methods and applicationsMagnetic Resonance Imaging, 1995
- High frequency volume coils for clinical NMR imaging and spectroscopyMagnetic Resonance in Medicine, 1994
- Mapping of the Radiofrequency FieldJournal of Magnetic Resonance, Series A, 1993
- Radiofrequency map of an NMR coil by imagingMagnetic Resonance Imaging, 1993
- Comparison of linear and circular polarization for magnetic resonance imagingJournal of Magnetic Resonance (1969), 1985