Respiratory flow integrator with improved stability
- 1 April 1978
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 44 (4) , 619-622
- https://doi.org/10.1152/jappl.1978.44.4.619
Abstract
A circuit was designed for integrating respiratory flow signals over time to produce volume signals. The circuit involves a simple integrator and three compensation subcircuits which control integrator drift caused by ambient temperature changes, flow signal base-line errors, and random signal errors. A functional unit built according to the circuit design requires initial adjustment, but no further adjustment before or during sustained use. Integration errors are typically less than 1%. The only flow signal requirements are that they be bidirectional and within the human physiological frequency range.This publication has 2 references indexed in Scilit: