A High Sensitivity EPR Technique for Alanine Dosimetry
- 1 May 1998
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 77 (1) , 43-49
- https://doi.org/10.1093/oxfordjournals.rpd.a032293
Abstract
Uncertainties of ± 5 mGy were achieved in the measurement of alanine dosemeters using optimised EPR parameters, instrumentation, spectral manipulation and subtraction techniques. Modulation amplitude and microwave power were adjusted to combine resonances of two neighbouring alanine signals. Instrumental variations were minimised by combining and subtracting pre- and post-measurement spectra of the empty EPR tube. A spectrum of the native signal of non-dosed alanine was generated from a single batch of dosemeters and subtracted from spectra of the irradiated dosemeters, also from the same batch. Field alignment was adjusted with the use of an in-cavity Mn ++ standard. A constant rotation goniometer was used to eliminate anisotropies in the EPR tube and alanine samples. Finally, digital filters were applied to the resulting spectra.Keywords
This publication has 0 references indexed in Scilit: