Measurement of Tidal Volume during High Frequency Ventilation by Impedance Plethysmography
- 1 March 1988
- journal article
- research article
- Published by Springer Nature in Pediatric Research
- Vol. 23 (3) , 253-256
- https://doi.org/10.1203/00006450-198803000-00003
Abstract
Electrical impedance plethysmography was evaluated in lambs as a method of measuring tidal volume (VT). Over tidal volumes ranging from 15 to 414% of estimated dead space, and frequencies of 300 to 1000 breaths/min, correlation between VT measured by the impedance technique and VT measured by whole-body plethysmography was 0.98. Above 600 breaths/min, the correlation between the two methods was 0.94. Independent calibration of the impedance technique using a pneumotachograph at conventional rates of ventilation yielded absolute values of VT which closely corresponded to values obtained with the whole-body plethysmograph (slope = 1.05, intercept = 1.4 ml, r = 0.99). These results support the potential utility of impedance plethysmography in clinical applications of high frequency ventilation.This publication has 3 references indexed in Scilit:
- Measurement of Tidal Volume during High- Frequency Jet Ventilation1Pediatric Research, 1985