Performance evaluation of a large axial field-of-view PET scanner: SET-2400W
- 1 December 1997
- journal article
- research article
- Published by Springer Nature in Annals of Nuclear Medicine
- Vol. 11 (4) , 307-313
- https://doi.org/10.1007/bf03165298
Abstract
The SET-2400W is a newly designed whole-body PET scanner with a large axial field of view (20 cm). Its physical performance was investigated and evaluated. The scanner consists of four rings of 112 BGO detector units (22.8 mm in-plane × 50 mm axial × 30 mm depth). Each detector unit has a 6 (in-plane) × 8 (axial) matrix of BGO crystals coupled to two dual photomultiplier tubes. They are arranged in 32 rings giving 63 two-dimensional image planes. Sensitivity for a 20-cm cylindrical phantom was 6.1 kcps/kBq/m/ (224 kcps/μCi/ml) in the 2D clinical mode, and to 48.6 kcps/kBq/ ml (1.8 Mcps/μCi/ml) in the 3D mode after scatter correction. In-plane spatial resolution was 3.9 mm FWHM at the center of the field-of-view, and 4.4 mm FWHM tangentially, and 5.4 mm FWHM radially at 100 mm from the center. Average axial resolution was 4.5 mm FWHM at the center and 5.8 mm FWHM at a radial position 100 mm from the center. Average scatter fraction was 8% for the 2D mode and 40% for the 3D mode. The maximum count rate was 230 kcps in the 2D mode and 350 kcps in the 3D mode. Clinical images demonstrate the utility of an enlarged axial field-of-view scanner in brain study and whole-body PET imaging.Keywords
This publication has 14 references indexed in Scilit:
- A New PET Camera for Noninvasive Quantitation of Physiological Functional Parametric ImagesPublished by Elsevier ,1996
- Development of a 2-Dimensional Gamma Ray Position Sensitive Detector for PET.RADIOISOTOPES, 1996
- Singles transmission in volume-imaging PET with a137Cs sourcePhysics in Medicine & Biology, 1995
- The ECAT EXACT HRJournal of Computer Assisted Tomography, 1994
- Performance Evaluation of the Positron Scanner ECAT EXACTJournal of Computer Assisted Tomography, 1992
- Fully three-dimensional reconstruction for a PET camera with retractable septaIEEE Transactions on Medical Imaging, 1991
- 3D PET using a Conventional Multislice Tomograph without SeptaJournal of Computer Assisted Tomography, 1991
- Implementation of three-dimensional image reconstruction for multi-ring positron tomographsPhysics in Medicine & Biology, 1990
- Design and evaluation of HEADTOME-IV, a whole-body positron emission tomographIEEE Transactions on Nuclear Science, 1989
- Fully-three-dimensional positron emission tomographyPhysics in Medicine & Biology, 1980