Image Reconstruction for a Rotating Positron Tomograph
- 1 January 1983
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 30 (1) , 594-600
- https://doi.org/10.1109/tns.1983.4332340
Abstract
A high-resolution positron tomograph consisting of two high-density avalanche chambers mounted on a rotating gantry has been installed in the Nuclear Medicine Department of the Cantonal Hospital in Geneva. Positron annihilation data are collected from six detector positions and back-projected in real time to form a single three-dimensional image. Up to 16 transverse or longitudinal sections may be imaged simultaneously. The derivation of the appropriate deconvolution filter is described in this paper, and the first images taken in the transverse mode are presented. It is shown that quantitative information can be obtained from such images.Keywords
This publication has 5 references indexed in Scilit:
- A general method for three-dimensional filter computation (positron emission tomography application)Physics in Medicine & Biology, 1983
- The High-Density Avalanche Chamber for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1983
- Filters for three-dimensional limited-angle tomographyPhysics in Medicine & Biology, 1981
- Fully-three-dimensional positron emission tomographyPhysics in Medicine & Biology, 1980
- Three-dimensional imaging in the positron camera using Fourier techniquesPhysics in Medicine & Biology, 1977