Prolonged Partial Venoarterial Bypass:Physiologic, Biochemical, and Hematologic Responses
- 1 December 1974
- journal article
- Published by Wolters Kluwer Health in Annals of Surgery
- Vol. 180 (6) , 850-856
- https://doi.org/10.1097/00000658-197412000-00010
Abstract
Prolonged veno arterial extracorporeal bypass (50-80% of cardiac output) was evaluated in normal, unanesthetized sheep. The evaluation protocol included serial measurements of hemodynamics, pulmonary and renal function, serum enzymes to detect organ damage, and detailed studies of coagulation and platelets. Blood exposure surfaces were primarily polyvinyl chloride and silicone rubber. Gas interfaces were carefully excluded with the exception of four experiments utilizing bubble oxygenators. Heparin dose was titrated to maintain activated clotting to two to three times baseline. Characteristics of 48 hour uncomplicated extracorporeal circulation in 8 sheep included normal hemodynamics, mild respiratory alkolosis negligable hemolysis, slight gradual increase in heart, liver, and muscle enzymes. The most significant changes occurred in coagulation and platelets characterized by an initial reduction in coagulation factors with a continued reduction in platelet count and return to normal clotting factors during extracorporeal circulation. This is followed by a two times normal increase in platelets and fibrinogen following extracorporeal circulation.Keywords
This publication has 17 references indexed in Scilit:
- Extracorporeal oxygenation for acute respiratory insufficiency: clinical experience.1973
- Prolonged use of a membrane oxygenator in air-breathing and hypoxic lambs. I. Venovenous bypass.1973
- Effects of Dexamethasone and Chlorpromazine on Skeletal Muscle Surface Hydrogen Ion Concentration during Cardiopulmonary BypassAnnals of Surgery, 1972
- Prolonged Extracorporeal Oxygenation for Acute Post-Traumatic Respiratory Failure (Shock-Lung Syndrome)New England Journal of Medicine, 1972
- Blood coagulation values of sheep.1971
- A comparison of the Landé-Edwards, the Peirce, and the General Electric-Peirce membrane lungs.1970
- SPLIT PRODUCTS OF FIBRIN IN THE SERUM OF NEWBORNSPediatrics, 1969
- A toroidal flow membrane oxygenator: four day partial bypass in dogs.1969
- Platelet Aggregation and the Availability of Platelet Factor 3British Journal of Haematology, 1966
- Changes in serum enzyme levels accompanying cardiac surgery with extracorporeal circulationJournal of Clinical Pathology, 1966