Calibration of the Malvern particle sizer
- 15 July 1984
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 23 (14) , 2415-2419
- https://doi.org/10.1364/ao.23.002415
Abstract
A relatively simple technique has been developed to calibrate particle- and droplet-sizing instruments manufactured by Malvern and other similar Fraunhofer diffraction particle-sizing instruments. Measurements of standard reticles using an instrument calibrated in this manner demonstrate the effectiveness of the procedure. In addition, guidelines are presented to avoid errors due to vignetting of the scattered light signal.Keywords
This publication has 1 reference indexed in Scilit:
- Droplet Size Distribution in SpraysIndustrial & Engineering Chemistry, 1951