A Comparison of Peritest Automated Perimetry and Goldmann Perimetry
- 1 March 1985
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 103 (3) , 397-403
- https://doi.org/10.1001/archopht.1985.01050030093030
Abstract
• Both Peritest perimetry and Goldmann visual field testing were performed on three groups of eyes. Eighty-one eyes had glaucoma with elevated intraocular pressures and abnormal visual fields determined by Goldmann perimetry. There were 47 eyes with suspected glaucoma, based on elevated IOPs and normal Goldmann visual fields. Nineteen eyes were normal. All visual fields were evaluated in masked fashion, and the results were compared. The Peritest had a high sensitivity in both glaucomatous (98.8%) and normal (94.7%) eyes. One half of the eyes with suspected glaucoma had abnormal Peritest results but normal Goldmann visual fields. These eyes were examined prospectively with color stereo discs and monochromatic nerve fiber layer photographs. One half of the eyes with suspected glaucoma and abnormal Peritest results also had photographic evidence suggesting optic nerve damage. Thus, early field defects detected by the Peritest perimeter in eyes with suspected glaucoma are likely to be representative of glaucomatous damage.This publication has 15 references indexed in Scilit:
- Automated PerimetryPublished by Jaypee Brothers Medical Publishing ,2011
- Variation in Visual Field Measurements with an Automated PerimeterAmerican Journal of Ophthalmology, 1984
- Detecting Glaucomatous Damage with the Friedmann Analyzer Compared with the Goldmann Perimeter and Evaluation of Stereoscopic Photographs of the Optic DiskAmerican Journal of Ophthalmology, 1983
- The Peritest, a new automatic and semi-automatic perimeterInternational Ophthalmology, 1982
- Variability of Static Visual Threshold Responses in Patients With Elevated IOPsArchives of Ophthalmology (1950), 1982
- Clinical Experience with the Fieldmaster PerimeterAmerican Journal of Ophthalmology, 1982
- A Clinical Comparison of Three Computerized Automatic Perimeters in the Detection of Glaucoma DefectsArchives of Ophthalmology (1950), 1981
- Computerized Profile Perimetry in GlaucomaArchives of Ophthalmology (1950), 1980
- Automatic (Octopus) and manual (Goldmann) perimetry in glaucomaAlbrecht von Graefes Archiv für Ophthalmologie, 1980
- Automated Suprathreshold Static PerimetryAmerican Journal of Ophthalmology, 1980