An adaptive breath sampler for use with human subjects with an impaired respiratory function
- 6 November 2006
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 132 (2) , 153-163
- https://doi.org/10.1039/b608608j
Abstract
An adaptive sampler for collecting 2.5 dm3 samples of exhaled air from human subjects with an impaired respiratory function is described. Pressure in the upper respiratory tract is continuously monitored and the data used to control an automated system to collect select portions of the expired breathing cycle onto a mixed bed Tenax™ and Carbotrap™ adsorbent trap for analysis by GC-MS. The sampling approach is intended for use in metabolomic profiling of volatiles in human breath at concentrations greater than µg m–3. The importance of experimental reproducibility in metabolomic data is emphasised and consequently a high purity air supply is used to maintain a stable exogenous volatile organic compound profile at concentrations in the range 5 to 30 µg m–3. The results of a 90 day stability study showed that exogenous VOCs were maintained at significantly lower levels (40 times lower for isopropyl alcohol) and with significantly higher reproducibility (80 times lower standard deviation for isopropyl alcohol) than would have been be the case if ambient air had been used. The sampling system was evaluated with healthy controls alongside subjects with chronic obstructive pulmonary disease. Subjects were able to breathe normally with control subjects observed to breathe at a rate of 9 to 17 breaths per minute, compared to 16 to 30 breaths per minute for subjects with COPD. This study presents, for the first time, observations and estimates of intra-subject breath sample reproducibility from human subjects. These reproducibility studies indicated that VOCs in exhaled breath exhibit a variety of dynamic behaviours, with some species recovered with a RSD <30%, while other species were observed to have significantly more variable concentrations, 30 to 130% RSD. The approach was also demonstrated to reliably differentiate the differences in the VOC profiles between alveolar and dead space air.Keywords
This publication has 26 references indexed in Scilit:
- Accuracy of breath tests using low doses of 13C-urea to diagnose Helicobacter pylori infection: a randomised controlled trialGut, 2006
- Solid phase microextraction for analysis of alkanes and aromatic hydrocarbons in human breathJournal of Chromatography B, 2005
- Preliminary investigation of using volatile organic compounds from human expired air, blood and urine for locating entrapped people in earthquakesJournal of Chromatography B, 2005
- Determination of acetone, hexanal and heptanal in blood samples by derivatization with pentafluorobenzyl hydroxylamine followed by headspace single-drop microextraction and gas chromatography–mass spectrometryAnalytica Chimica Acta, 2005
- Effects of ventilation on the collection of exhaled breath in humansJournal of Applied Physiology, 2004
- Personal volatile organic compound (VOC) exposure of children attending elementary schools adjacent to industrial complexAtmospheric Environment, 2004
- Solid-Phase Microextraction for the Analysis of Human BreathAnalytical Chemistry, 1997
- Effect of relative humidity on the performance of a platinum-lead denuder for the adsorption of nitrobenzeneAnalytica Chimica Acta, 1993
- Automatic end-expiratory air sampling device for breath hydrogen test in infantsThe Lancet, 1991
- Quantitative Analysis of Urine Vapor and Breath by Gas-Liquid Partition ChromatographyProceedings of the National Academy of Sciences, 1971