Corrosion testing of stents: A novel fixture to hold entire device in deployed form and finish
- 1 January 1999
- journal article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 48 (6) , 829-832
- https://doi.org/10.1002/(sici)1097-4636(1999)48:6<829::aid-jbm10>3.0.co;2-#
Abstract
Stainless steel and nickel-titanium alloy stents for cardiac, biliary, and prostatic clinical applications may be susceptible to loss of mechanical integrity/failure due to their small size and unique geometry. Because their corrosion behavior is a critical aspect of their biocompatibility, it is necessary to test such devices for corrosion behavior using a final deployed form and finish condition. Welding wire leads or holding with metallic clips can result in surface variability or damage to the stents. A novel fixture to hold stents for corrosion testing was designed and evaluated in this study. This fixture design minimally influenced the properties of stents, and also facilitated easy removal post-testing for other types of characterization.Keywords
This publication has 3 references indexed in Scilit:
- Corrosion resistance tests on NiTi shape memory alloyBiomaterials, 1996
- Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants.Journal of Bone and Joint Surgery, 1994