Clinical utility of a colorimetric end-tidal CO2 detector in cardiopulmonary resuscitation and emergency intubation
- 1 October 1991
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 7 (4) , 289-293
- https://doi.org/10.1007/bf01619347
Abstract
The purposes of this study were to evaluate the clinical utility of a colorimetric end-tidal CO2 (ETCO2) detector in confirming proper endotracheal intubation in patients requiring emergency intubation, to determine if this new device can be used as an adjunct to judge the effectiveness of cardiopulmonary resuscitation (CPR), and to determine whether the device can predict successful resuscitation from cardiopulmonary arrest. We studied prospectively 110 patients requiring emergency intubation for either respiratory distress (53 patients) or cardiopulmonary arrest (57 patients) by recording the color range of the indicator after the initial intubation. In patients who suffered a cardiopulmonary arrest, the color range was also recorded during CPR after the endotracheal tube was confirmed to be in the tracheal position and perfusion optimized, and at the moment CPR was stopped. The ETCO2 detector was 100% specific for correct endotracheal intubation in all patients. It was also highly sensitive (0.98) for correct endotracheal intubation in patients with respiratory distress. However, it was not sensitive (0.62) in patients with cardiopulmonary arrest and low perfusion. The sensitivity improved (0.88) when we used the ETCO2 range obtained after attempts to increase perfusion. A low ETCO2 color range in 19 patients undergoing CPR was interpreted as low cardiac output and prompted the physicians to attempt to increase perfusion. Of the patients who underwent CPR, no patient whose ETCO2 level remained less than 2% was successfully resuscitated. Those patients who had an ETCO2 level ≥2% had a significantly higher incidence of successful resuscitation. We conclude that the colorimetric ETCO2 detector is reliable and provides reassurance of correct endotracheal tube placement in patients requiring emergency intubation for respiratory distress. This device helps identify patients with low perfusion during CPR and is a useful prognostic indicator of successful short-term resuscitation.Keywords
This publication has 24 references indexed in Scilit:
- END TIDAL CO2 AND HIGH DOSE EPINEPHRINE DURING CARDIAC ARREST IN HUMANSCritical Care Medicine, 1990
- DETECTION OF ESOPHAGEAL INTUBATION WITH A NEW CHEMICAL MONITOR OF END-TIDAL CARBON DIOXIDECritical Care Medicine, 1990
- Prediction of outcome of cardiopulmonary resuscitation from end-tidal carbon dioxide concentrationCritical Care Medicine, 1990
- DOES IT TAKE LONGER TIME TO INDUCE CHILDREN IN THE PEDIATRIC INDUCTION ROOM IN THE PARENTʼS PRESENCE?Anesthesia & Analgesia, 1990
- Capnographic waveforms in esophageal intubation: Effect of carbonated beveragesAnnals of Emergency Medicine, 1989
- Early detection of inadvertent oesophageal intubation: pulse oximetry vs. capnographyActa Anaesthesiologica Scandinavica, 1989
- End-Tidal Carbon Dioxide Concentration during Cardiopulmonary ResuscitationNew England Journal of Medicine, 1988
- Esophageal IntubationAnesthesia & Analgesia, 1986
- Cardiac output and end-tidal carbon dioxideCritical Care Medicine, 1985
- Expired PCO2 as a prognostic indicator of successful resuscitation from cardiac arrestAnnals of Emergency Medicine, 1985