Calorimetric Determination of Absorbed Dose and G Fe +++ of the Fricke Dosimeter with 10-Mev and 20-Mev Electrons
- 1 November 1963
- journal article
- research article
- Published by JSTOR in Radiation Research
- Vol. 20 (3) , 423-+
- https://doi.org/10.2307/3571372
Abstract
The construction and operational details of a local absorbed dose microcalorimeter are presented. The calorimeter measures the absorbed dose rate at any desired point within an irradiated medium. Measurement is specific to a sufficiently limited region (2.0 cm in diameter, 0.3 cm deep) so that it is essentially at a point and no integration or averaging process is necessary to qualify either the absorbed dose, or its rate. The system is sufficiently sensitive to provide about 1% accuracy at dose rates of 50 rads/min. GFe+++ at 10 Mev was found to be 15.32 [plus or minus] 0.34; at 20 Mev, 15.17 [plus or minus] 0.23.Keywords