The use of a holographic rejection filter has been demonstrated to improve the sensitivity of a commercial Raman microprobe by a factor of ∼3.5, allowing shorter data accumulation times or a reduction of the laser intensity at the sample. The filter has been tested with two very different solid samples; the first (TiO2 powder) gives a strong diffusely scattered laser beam along with the Raman scatter. The second (silicon) gives a strong specular reflection of the laser beam, again severely testing the laser rejection capabilities of the filter. Bands down to 200 cm−1 shift were transmitted by the filter without attenuation but with satisfactory rejection of the laser line. This approach provides a simple way of enhancing the performance of a commercial microprobe at low cost.Keywords
This publication has 4 references indexed in Scilit:
- Characterization of Holographic Band-Reject Filters Designed for Raman SpectroscopyApplied Spectroscopy, 1991
- Compact Raman instrumentation for process and environmental monitoringPublished by SPIE-Intl Soc Optical Eng ,1991
- The Utilization of a Holographic Bragg Diffraction Filter for Rayleigh Line Rejection in Raman SpectroscopyApplied Spectroscopy, 1990
- Novel applications of Raman microscopyPhilosophical Transactions of the Royal Society of London. Series A, Mathematical and Physical Sciences, 1986