Measured overall perturbation factors at depths greater than for ionization chambers in electron beams
- 12 February 1999
- journal article
- Published by Wiley in Medical Physics
- Vol. 26 (2) , 208-213
- https://doi.org/10.1118/1.598506
Abstract
In electron beam dosimetry the perturbation effect in the medium by the ionization chamber cavity is accounted for by introducing a replacement correction factor, Another perturbation correction factor, denoted as is due to the materials of the walls of the parallel‐plate chamber differing from the phantom material. Because of the difficulties in separating these two components, we measure the overall perturbation factor, A distinct advantage of parallel‐plate ionization chambers over cylindrical chambers is that has been shown to be close to unity at the standard calibration depth, However, for many dosimetry applications it is necessary to know the overall perturbation factor at depths greater than We measured the overall perturbation factor at depths greater than (approximating the 95%, 90% and 50% depth dose) for a Farmer‐type cylindrical ionization chamber and three parallel‐plate ionization chambers. We assumed that for the NACP chamber is unity at these measurement depths. The depth dependence for the other chambers was then measured relative to the NACP chamber. The mean energy at depth, and percentage depth dose gradient ranges studied were 1.9–18.5 MeV and 0 to 4.5%/mm, respectively. For the other two parallel‐plate chambers, we find to be unity at depths where the percent depth dose is greater than 90%, but it deviates from unity at deeper depths, where the dose gradients exceed about 2.5%/mm. For the cylindrical chamber, values at depths greater than were found to be in good agreement with those in TG 21, where the energy at depth, is used to evaluateKeywords
This publication has 12 references indexed in Scilit:
- R50 as a beam quality specifier for selecting stopping‐power ratios and reference depths for electron dosimetryMedical Physics, 1996
- The calibration and use of plane‐parallel ionization chambers for dosimetry of electron beams: An extension of the 1983 AAPM protocol report of AAPM Radiation Therapy Committee Task Group No. 39Medical Physics, 1994
- Verification of absorbed doses determined with thimble and parallel-plate ionization chambers in clinical electron beams using ferrous sulphate dosimetryMedical Physics, 1994
- Experimental values for P wall, x and P repl, E for five parallel-plate, ion chambers-A new analysis of previously published dataMedical Physics, 1992
- Perturbation correction factors for some ionization chambers commonly applied in electron beamsPhysics in Medicine & Biology, 1991
- Measurement of the replacement correction factor for parallel-plate chambers in electron fieldsMedical Physics, 1991
- Implication of electron backscattering for electron dosimetryPhysics in Medicine & Biology, 1991
- Electron backscatter corrections for parallel‐plate chambersMedical Physics, 1988
- Calibration and use of Plane-Parallel Ionization Chambers for the Determination of Absorbed Dose in Electron BeamsActa Radiologica: Oncology, 1981
- Einfluß der Vielfachstreuung von Elektronen auf die Ionisation in gasgefüllten HohlräumenRadiation and Environmental Biophysics, 1968