Speer carbon resistance thermometer magnetoresistance effect
- 1 August 1977
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 48 (8) , 1090-1091
- https://doi.org/10.1063/1.1135190
Abstract
Speer carbon resistors (100 Ω, grade 1002) have been calibrated in the range 5–50 mK using a 60Co Co nuclear orientation thermometer. Negative longitudinal magnetoresistance effects have been observed up to 2.5 T.Keywords
This publication has 4 references indexed in Scilit:
- Reproducibility of carbon resistors used as thermometers below 1°KPublished by Elsevier ,2002
- Some observations on resistance thermometry below 1 KReview of Scientific Instruments, 1976
- Thermal resistance in Speer carbon resistors below 1 KCryogenics, 1974
- Speer Carbon Resistors as Thermometers for Use Below 1°KReview of Scientific Instruments, 1964