Sources of Error in the Determination of Output Voltage of Pulse Generators by Pacemaker System Analyzers
- 1 September 1980
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 3 (5) , 585-596
- https://doi.org/10.1111/j.1540-8159.1980.tb05278.x
Abstract
The use of threshold or pacemaker system analyzers with widely different characteristics has introduced potential sources of error in the determination of the output voltage of pulse generators. This is further compounded by the availability of pulse generators with diverse waveform configurations and programmability capabilities. Because of this non‐uniformity, the physician must have some rudimentary knowledge of waveform characteristics and appreciate the limitations of threshold or pacemaker system analyzers to avoid the unnecessary replacement of normally‐functioning pulse generators.Keywords
This publication has 3 references indexed in Scilit:
- Social Mobility in Industrial Labor MarketsActa Sociologica, 1979
- Definitions and Glossary (Part I)Pacing and Clinical Electrophysiology, 1979
- Techniques and Significance of Threshold Measurement for Cardiac PacingChest, 1976