An intravascular oxygen tension monitoring device used in myocutaneous transplants: A preliminary report
- 1 January 1993
- journal article
- research article
- Published by Wiley in Microsurgery
- Vol. 14 (5) , 304-309
- https://doi.org/10.1002/micr.1920140503
Abstract
An intravascular oxygen tension (PO2) sensor (Continucath 1000) was used to assess flap viability in microsurgically revascularized free flaps for 36 hours postoperatively. In vitro, long term stability, linearity, and response time of the PO2 sensor proved to be excellent. In 14 patients with sound flaps tissue PO2 values were found to be between 15 and 35 mm Hg. In three cases a failing flap needed reexploration. In these cases the PO2 sensor was accurate in detecting circulatory failure while in other cases it traced a number of physiological changes. The device holds promise for free flap monitoring but further studies are indicated. © 1993 Wiley-Liss Inc.Keywords
This publication has 8 references indexed in Scilit:
- Differential Effects of Venous Stasis and Arterial Insufficiency on Tissue Oxygenation in Myocutaneous Island FlapsPlastic and Reconstructive Surgery, 1992
- Tissue Oxygen Tension Measurement for Monitoring Musculocutaneous and Cutaneous FlapsScandinavian Journal of Plastic and Reconstructive Surgery and Hand Surgery, 1990
- Variations in Flap Blood Flow and Tissue Po2Annals of Plastic Surgery, 1988
- Assessment of perfusion in postoperative patients using tissue oxygen measurementsBritish Journal of Surgery, 1987
- Local Heat Increases Blood Flow and Oxygen Tension in WoundsArchives of Surgery, 1987
- Continuous direct tissue oxygen tension measurement by a new method using an implantable silastic tonometer and oxygen polarographyThe American Journal of Surgery, 1983
- Direct Measurement of Wound and Tissue Oxygen Tension in Postoperative PatientsAnnals of Surgery, 1983
- Measurement of oxygen tensionCritical Care Medicine, 1981