A Model of the Patient-Artificial Kidney System
- 1 July 1971
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 18 (4) , 257-264
- https://doi.org/10.1109/tbme.1971.4502848
Abstract
A mathematical analysis is made of the changes in body fluid volume and solute composition that occur during hemodialysis. The analysis considers the movement of water and solutes across the dialyzer membrane and between the intracellular and extracellular compartments. Dialyzer operating variables such as dialysate addition rate, solute dialysances, and ultrafiltration rate are included. The output of the mathematical model agrees well with data obtained during clinical dialysis. The results of the analysis indicate that the intracellular volume changes that would be expected during dialysis as a result of osmotic gradients are very small. The effects on solute removal of varying dialysate addition rate, ultrafiltration rate, and dialysance are analyzed.Keywords
This publication has 5 references indexed in Scilit:
- Production of Urea Nitrogen and Creatinine in Chronic Azotemia And Effect of Hemodialysis.Experimental Biology and Medicine, 1966
- ANALOG SIMULATION OF THE PATIENT-ARTIFICIAL KIDNEY SYSTEMAsaio Journal, 1965
- ELECTROENCEPHALOGRAPHIC CHANGES DURING HÆMODIALYSISThe Lancet, 1963
- Urea transport in the central nervous systemAmerican Journal of Physiology-Legacy Content, 1962