Dual-energy electronic scanning-slit fluorography for the determination of vertebral bone mineral content
- 1 July 1987
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 14 (4) , 562-566
- https://doi.org/10.1118/1.596068
Abstract
A technique for performing accurate bone mineral content (BMC) determination, using dual-energy scanning-slit fluorography, has been developed. X-ray scatter and veiling glare are suppressed using a multiple scanning-slit device interfaced to a conventional digital substraction angiography system. The technique employs a pair of calibration step wedges, one tissue and one bone equivalent. The proper use of the step wedges obviates the need for accurate kVp calibration or absolute calibration of the system''s response to BMC. Phantom studies intended to simulate vertebral BMC measurement indicate that the technique is capable of measuring BMC with less than 30 mg/cm2 inaccuracy over a wide range of BMC.Funding Information
- National Cancer Institute (5-R01-CA32854)
This publication has 7 references indexed in Scilit:
- Scatter rejection by electronic collimationMedical Physics, 1986
- Dual-Energy Projection Radiography in the Evaluation of Femoral Neck Strength, Density, and MineralizationInvestigative Radiology, 1985
- Precise Measurement of Vertebral Mineral Content Using Computed TomographyJournal of Computer Assisted Tomography, 1980
- Assessment of bone mass in the radius by computed tomographyAmerican Journal of Roentgenology, 1980
- Epoxy resin based tissue substitutesThe British Journal of Radiology, 1977
- NEW TECHNIC FOR RADIOGRAPHIC ASSESSMENT OF VERTEBRAL DENSITY1964
- Measurement of Bone Mineral in vivo: An Improved MethodScience, 1963