Performance characteristics of a new photometer with a moving filter tape for luminescence assay
- 1 December 1977
- journal article
- research article
- Published by American Society for Microbiology in Applied and Environmental Microbiology
- Vol. 34 (6) , 720-724
- https://doi.org/10.1128/aem.34.6.720-724.1977
Abstract
The performance characteristics of a new photometer that incorporated a reaction system consisting of a movable filtration tape have been studied for use with the firefly luciferase assay for adenosine 5'-triphosphate. Precision, linearity, and sensitivity are given for measuring a graded series of constant-light emitters, concentrations of adenosine 5'-triphosphate, and washed bacterial cells. Precision ranged from 1 to 10% coefficient of variation. The coefficient of correlation was 0.9985 for 7 X 10(8) to 7 X 10(3) bacteria per ml; however, the accuracy decreased at the low levels. The sample processing time was 2 min for a 1-ml sample. This instrument shows potential for many applications in quantitating small numbers of organisms, e.g., pollution, water monitoring, and clinical infection detection.This publication has 5 references indexed in Scilit:
- MICROASSAY OF ADENOSINE‐3′,5′‐MONOPHOSPHATE (CYCLIC AMP) IN BRAIN AND OTHER TISSUES BY THE LUCIFERIN‐LUCIFERASE SYSTEM1Journal of Neurochemistry, 1971
- MECHANISM OF BIOLUMINESCENCE, CHEMI‐LUMINESCENCE AND ENZYME FUNCTION IN THE OXIDATION OF FIREFLY LUCIFERIN*,†Photochemistry and Photobiology, 1969
- The decarboxylation of luciferin in firefly bioluminescenceBiochemical and Biophysical Research Communications, 1968
- The ATP pool in Escherichia coli. I. Measurement of the pool using a modified luciferase assayBiochimica et Biophysica Acta (BBA) - Bioenergetics, 1967