Matrix-Assisted Laser Desorption/Ionization Coupled with Quadrupole/Orthogonal Acceleration Time-of-Flight Mass Spectrometry for Protein Discovery, Identification, and Structural Analysis
- 14 March 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (8) , 1707-1720
- https://doi.org/10.1021/ac0011080
Abstract
The design and operation of a novel UV-MALDI ionization source on a commercial QqoaTOF mass spectrometer (Applied Biosystem/MDS Sciex QSTAR Pulsar) is described. Samples are loaded on a 96-well target plate, the movement of which is under software control and can be readily automated. Unlike conventional high-energy MALDI-TOF, the ions are produced with low energies (5 − 10 eV) in a region of relatively low vacuum (8 mTorr). Thus, they are cooled by extensive low-energy collisions before selection in the quadrupole mass analyzer (Q 1 ), potentially giving a quasi-continuous ion beam ideally suited to the oaTOF used for mass analysis of the fragment ions, although ion yields from individual laser shots may vary widely. Ion dissociation is induced by collisions with argon in an rf-only quadrupole cell, giving typical low-energy CID spectra for protonated peptide ions. Ions separated in the oaTOF are registered by a four-anode detector and time-to-digital converter and accumulated in “bins” that are 625 ps wide. Peak shapes depend upon the number of ion counts in adjacent bins. As expected, the accuracy of mass measurement is shown to be dependent upon the number of ions recorded for a particular peak. With internal calibration, mass accuracy better than 10 ppm is attainable for peaks that contain sufficient ions to give well-defined Gaussian profiles. By virtue of its high resolution, capability for accurate mass measurements, and sensitivity in the low-femtomole range, this instrument is ideally suited to protein identification for proteomic applications by generation of peptide tags, manual sequence interpretation, identification of modifications such as phosphorylation, and protein structural elucidation. Unlike the multiply charged ions typical of electrospray ionization, the singly charged MALDI-generated peptide ions show a linear dependence of optimal collision energy upon molecular mass, which is advantageous for automated operation. It is shown that the novel pulsing technique of this instrument that increases the sensitivity for precursor ions scans is applicable to the identification of peptides labeled with isotope-coded affinity tags.Keywords
This publication has 43 references indexed in Scilit:
- Signaling—2000 and BeyondCell, 2000
- Protein sequence and structural studies employing matrix-assisted laser desorption ionization-high energy collision-induced dissociationInternational Journal of Mass Spectrometry and Ion Processes, 1997
- High Sensitivity Collisionally-activated Decomposition Tandem Mass Spectrometry on a Novel Quadrupole/Orthogonal-acceleration Time-of-flight Mass SpectrometerRapid Communications in Mass Spectrometry, 1996
- Matrix Dependence of Metastable Fragmentation of Glycoproteins in MALDI TOF Mass SpectrometryAnalytical Chemistry, 1995
- A combined magnetic sector–time‐of‐flight mass spectrometer for structural determination studies by tandem mass spectrometryRapid Communications in Mass Spectrometry, 1995
- Delayed extraction matrix‐assisted laser desorption time‐of‐flight mass spectrometryRapid Communications in Mass Spectrometry, 1995
- Error-Tolerant Identification of Peptides in Sequence Databases by Peptide Sequence TagsAnalytical Chemistry, 1994
- Rapid identification of proteins by peptide-mass fingerprintingCurrent Biology, 1993
- Peptide sequencing by matrix‐assisted laser‐desorption mass spectrometryRapid Communications in Mass Spectrometry, 1992
- Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltonsAnalytical Chemistry, 1988