A simple theorem relating noise and patient dose in computed tomography
- 11 November 1999
- journal article
- Published by Wiley in Medical Physics
- Vol. 26 (11) , 2231-2234
- https://doi.org/10.1118/1.598778
Abstract
Presented is a simple model describing the dependence of image noise in computed tomography on the x-ray beam profile. The model is used to derive the x-ray profile which minimizes total image noise at constant integral patient dose. The profile may be produced with a bow-tie-type beam shaping filter. Results of the analysis are validated using a computer simulation of computed tomography (CT) acquisition and reconstruction.Keywords
This publication has 1 reference indexed in Scilit:
- A computer simulation of wavelet noise reduction in computed tomographyMedical Physics, 1999