Technical note: A novel impeller TAH using magnetic bearings for load reduction
- 1 January 2002
- journal article
- research article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 26 (5) , 214-216
- https://doi.org/10.1080/03091900210156805
Abstract
A novel impeller TAH (total artificial heart), i.e. bi-ventricular assist impeller pumps, has been developed. The device consists of a rotor with motor magnets and two impellers, a stator with motor coil and iron core, and two pump housings. In both sides of the rotor magnets, as well as the stator coil core, a pair of magnetic bearings was devised to partly counteract the attractive forces between the rotor magnets and the stator coil core. This means the magnetic bearings are used for load reduction. On hydrodynamic testing, the two pumps both produced a flow rate as high as 6 l min(-1) and the left pump had a pressure head of 150 mm Hg, and that of the right pump was 50 mm Hg. The highest efficiency of the device, including the motor, the two pumps and the controller, reached 14.7%. The device, weighing 250 g, had a length of 80 mm and a diameter of 40 mm at its largest point. Currently in the world, this is a unique TAH, which is electrically powered and driven by a single motor and has only one moving part, can produce either pulsatile or non-pulsatile flow, both pumps eject flow synchronistically by pulsatile mode, and the volume equilibrium of the two pumps can be achieved automatically without the need for control.Keywords