In Vivo Measurement of Hepatic Iodine Concentration Using Fluorescent Excitation Analysis
- 1 March 1976
- journal article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 11 (2) , 134-137
- https://doi.org/10.1097/00004424-197603000-00008
Abstract
Hepatic iodine concentration was measured in the live dog by external use of fluorescent excitation analysis. The number of characteristic photons produced by interaction of exciting radiation from an americium-241 source with iodine within the tissue is proportional to the tissue iodine concentrations. A correction is made for absorption of radiation by the abdominal wall and other tissues lying between the volume of liver being assayed and the detector collimator. The technique is applicable to the in vivo measurement of iodine concentrations from 0.5 to 40 mg/g. Accuracy of the technique is approximately +/- 10%, which is within the range of variation in iodine concentration at various sites within the liver. Radiation dose is low, and radiolabeled tracer compounds need not be used.Keywords