Abstract
The authors compared referee (senior author) microscopic counts, microscopic counts by several technologists, and counts obtained with two pattern-recognition leukocyte classifiers, (1) Larc® and (2) Hematrak®, and a cytochemical automated method for leukocyte counting, (3) Hemalog D®, using samples from (1) a random patient population, (2) a selected abnormal patient population, and (3) healthy individuals. All instruments showed good accuracy and flagged abnormal results for review. Variability in pattern-recognition counts was found to be due mainly to the distribution of the cells on prepared blood smears. The Larc classifier was found to be very sensitive to minor alterations in the cytoplasm or nucleus, and rejected a number of slides. The Hemalog D showed the greatest precision. The method of preparation of slides would be a major decision factor in selecting one pattern-recognition instrument over another.Keywords
This publication has 1 reference indexed in Scilit:
- An Automatic Leukocyte Analyzer: Validity of Its ResultsAmerican Journal of Clinical Pathology, 1977