The measurement of energy imparted to patients during diagnostic x-ray examinations using the Diamentor exposure-area product meter
- 1 October 1984
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 29 (10) , 1199-1208
- https://doi.org/10.1088/0031-9155/29/10/003
Abstract
The mean energy imparted to the patient has been proposed as a practical quantity which may be useful for assessing somatic risk in X-ray diagnostic radiology. A technique is described for estimating energy imparted during most common X-ray examinations of the trunk or complete head from a measurement of exposure-area product conveniently provided by the Diamentor transmission ionization chamber. The necessary energy imparted calibration for the Diamentor was deduced from consideration of incident spectral energy fluence and Monte Carlo calculations of the fraction of total beam energy imparted to mathematical phantoms representing the patient. The relationship between energy imparted and exposure-area product depends primarily on the applied potential and total filtration of the X-ray beam and, to a lesser extent, on voltage waveform and X-ray target angle. Direct measurements of energy imparted to an Alderson Rando phantom for a range of irradiation conditions provided an excellent verification of the technique developed. Using the energy imparted per exposure-area product factors presented, overall uncertainties of less than .+-.15% and .+-.20% should be possible for measurements of energy imparted during examinations of the trunk and head, respectively, using the Diamentor.This publication has 7 references indexed in Scilit:
- Comparison of methods for estimating the energy imparted to patients during diagnostic radiological examinationsPhysics in Medicine & Biology, 1983
- A re-evaluation of the 'saturated scatter' method for estimating the energy imparted to patients during diagnostic radiology examinationsPhysics in Medicine & Biology, 1983
- An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinationsPhysics in Medicine & Biology, 1982
- A reappraisal of the genetic consequences of diagnostic radiology in Great BritainThe British Journal of Radiology, 1981
- Tissue substitutes in experimental radiation physicsMedical Physics, 1978
- Photon Mass Attenuation and Mass Energy-Absorption Coefficients for H, C, N, O, Ar, and Seven Mixtures from 0.1 keV to 20 MeVRadiation Research, 1977
- Determination of Integral Absorbed Dose from Exposure MeasurementsActa Radiologica: Therapy, Physics, Biology, 1963