Randomised controlled study of clinical efficacy of spacer therapy in asthma with regard to electrostatic charge
Open Access
- 1 February 2001
- journal article
- clinical trial
- Published by BMJ in Archives of Disease in Childhood
- Vol. 84 (2) , 178-182
- https://doi.org/10.1136/adc.84.2.178
Abstract
BACKGROUND Inhalation therapy using a pressured metered dose inhaler (pMDI) and a spacer is frequently used in the treatment of airway disease in children. Several laboratory studies found a clear negative influence of electrostatic charge (ESC) on plastic spacers on the delivery of aerosol. AIMS To investigate whether ESC on plastic spacers could diminish bronchodilating responses to salbutamol. METHODS Ninety asthmatic children (aged 4–8 years) were randomised into three groups: metal Nebuchamber, plastic Volumatic, and plastic Aerochamber. The bronchodilating response was measured by the change in peak expiratory flow rate (PEF) after 100 μg and 400 μg salbutamol. Within the Volumatic and Aerochamber groups, a crossover comparison was made between electrostatic and non-electrostatic spacers. RESULTS We found no significant effect of ESC on the bronchodilating response to salbutamol with any of the doses in the Aerochamber and Volumatic groups. For the plastic spacers, the mean difference of the change in PEF after 100 μg salbutamol between non-electrostatic and electrostatic spacers was only +1.7% (95% CI −1.3% to 4.7%). After 400 μg salbutamol this was +1.9% (95% CI −1.4% to 5.1%). A comparable efficacy was found for the Nebuchamber, the Aerochamber, and Volumatic with respect to the change in PEF after 100 and 400 μg salbutamol. CONCLUSION This study showed no negative influence of ESC on plastic spacers with regard to clinical efficacy of a β2 agonist (salbutamol) in children with asthma. The metal Nebuchamber, plastic Aerochamber, and plastic Volumatic were equally effective.Keywords
This publication has 23 references indexed in Scilit:
- Washing plastic spacers in household detergent reduces electrostatic charge and greatly improves deliveryEuropean Respiratory Journal, 1999
- Early lung absorption profile of non‐CFC salbutamol via small and large volume plastic spacer devicesBritish Journal of Clinical Pharmacology, 1998
- The effects of static charge in spacer devices on glucocorticosteroid aerosol deposition in asthmatic patientsEuropean Respiratory Journal, 1998
- Use of budesonide in severe asthmatics aged 1-3 years.Archives of Disease in Childhood, 1993
- Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug.Thorax, 1993
- A comparison of six different ways of expressing the bronchodilating response in asthma and COPD; reproducibility and dependence of prebronchodilator FEV1European Respiratory Journal, 1992
- Comparison of terbutaline via the Nebuhaler and salbutamol via the Volumatic: theory and practiceEuropean Respiratory Journal, 1990
- A trial of metaproterenol by metered‐dose inhaler and two spacers in preschool asthmaticsPediatric Pulmonology, 1988
- Evaluation of Inhalation Aids of Metered Dose Inhalers in Asthmatic ChildrenChest, 1987
- The functional development of the respiratory system from the period of gestation to adulthood.Published by Elsevier ,1979