Sensitivity and Specificity of Standard and Rapid HIV‐Antibody Tests Evaluated by Seroconversion and Non‐Seroconversion Low‐Titre Panels
- 1 January 1997
- journal article
- Published by Wiley in Vox Sanguinis
- Vol. 72 (1) , 11-15
- https://doi.org/10.1046/j.1423-0410.1997.00011.x
Abstract
The aim of this study is to compare the relative sensitivity and specificity of commercial HIV-antibody assays using seroconversion, non-seroconversion panels, and negative blood donor samples.We evaluated the sensitivity of five standard ELISA HIV-antibody assays: Vironostika HIV Uni-Form II, Abbott recombinant HIV-1/HIV-2 third-generation EIA, Biotest Anti-HIV-1/-2 recombinant, Recombigen HIV-1/ HIV-2 EIA and Wellcozyme HIV 1 + 2 (VK54/55), and three rapid screening tests, Capillus HIV-1/HIV-2, Abott Test Pack HIV-1/HIV-2 third-generation EIA, and Sensy-Test HIV 1/2. All tests were assessed using four panels of plasma samples obtained from individuals who were seroconverting and a low-titre HIV-antibody panel of samples. Specificity of the standard screening tests was determined on 3.500 HIV-antibody-negative blood donor samples.There was no statistically significant difference in sensitivity between the five standard ELISA tests. One of these tests was significantly less specific than the others. The standard ELISA tests detected all the low-titre HIV-antibody-positive samples. Two of the rapid screening tests were significantly less sensitive on the seroconversion panels and all three tests failed to detect at least one of the positive samples in the low-titre panel.The additional risk of using one or other of the standard ELISA tests under review of not detecting all HIV-positive units of blood is not statistically significant. Using some of the rapid screening tests will, however, add a significant additional risk. A rapid screening test should therefore be adopted only after careful consideration of the effect of a possible lack of sensitivity on the safety of the blood supply.Keywords
This publication has 6 references indexed in Scilit:
- Sensitivity of HIV-antibody assays determined by seroconversion panelsAIDS, 1994
- Early detection of antibodies to HIV-1 by third-generation assaysThe Lancet, 1992
- Early detection of antibody to human immunodeficiency virus type 1 by using an antigen conjugate immunoassay correlates with the presence of immunoglobulin M antibodyJournal of Clinical Microbiology, 1992
- FOLLOWING UP INDETERMINATE HIV-1 WESTERN BLOTSThe Lancet, 1989
- Assessment of HIV‐1 screening test sensitivities using serially diluted positive sera can give misleading resultsTransfusion, 1989