FID‐based lung MRI at 0.5 T: Theoretical considerations and practical implications
- 1 April 1998
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 39 (4) , 666-672
- https://doi.org/10.1002/mrm.1910390423
Abstract
Pulse sequences based on FID signals and projection reconstruction (PR) were investigated for lung MRI at 0.5 T and evaluated for artifacts caused by: (1) k‐space mismapping due to either delay or distortion of the readout gradient waveform, (2) cardiac motion and pulsatile flow, and (3) respiratory motion. Nonstructured artifacts were described, simulated, and experimentally confirmed for the first time. Nonstructured artifacts did not impair the demonstration of structures of high signal‐to‐noise ratio (SNR) but generated quantitative errors in the image intensity analysis over the lung parenchyma. The use of FID‐based PR techniques for lung MRI is not justified at 0.5 T.Keywords
This publication has 16 references indexed in Scilit:
- Measurement of Lung Water Content and Pleural Pressure Gradient with Magnetic Resonance ImagingJournal of Thoracic Imaging, 1995
- A monitoring, feedback, and triggering system for reproducible breath‐hold MR imagingMagnetic Resonance in Medicine, 1993
- MR microscopy of the rat lung using projection reconstructionMagnetic Resonance in Medicine, 1993
- Magnetic resonance imaging of the lung parenchyma — Possible sequencesClinical Radiology, 1992
- Projection Reconstruction Techniques for Reduction of Motion Effects in MRIMagnetic Resonance in Medicine, 1992
- MR imaging of the lungs: value of short TE spin-echo pulse sequences.American Journal of Roentgenology, 1992
- Deblurring for non‐2D fourier transform magnetic resonance imagingMagnetic Resonance in Medicine, 1992
- Band-selective pulses without phase distortion. A simulated annealing approachJournal of Magnetic Resonance (1969), 1989
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988