Coronary angioplasty through 4 french diagnostic catheters

Abstract
In 50 consecutive patients subjected to coronary angioplasty immediately following a 4 French (F) diagnostic study, the technical feasibility and economical aspects of angioplasty through 4F catheters of 54 lesions were assessed. The patients were selected, but multiple, eccentric, and long lesions were not a priori excluded. 4F diagnostic catheters (Cordis), and fixed‐wire dilatation catheters (Ace, Scimed) were used in all cases. The procedure was successful in 43 lesions (80%) using 4F catheters. For 11 stenoses (20%), a change over to a larger French size was required. Two of these lesions could not be crossed with the balloon despite the larger sized guiding catheter. The final overall success rate was 96%, and there were no major complications. The use of diagnostic 4F catheters for angioplasty in these 50 patients resulted in the saving of 39 guiding catheters and 19 introducer sheaths. For 12 lesions (22%), an additional 4F catheter became necessary since the shape used for the diagnostic study was inadequate for angioplasty. In 7 cases, more than 1 balloon was used, but 5 of these balloon exchanges were independent of the use of 4F catheters. Three exchanges were performed through the 4F catheter (1 for need of a larger balloon to improve on an unsatisfactory angiographic result and 2 for a crimped guide wire tip of the Ace balloon). In the remaining 4, a larger catheter was used; in 2 of them, angioplasty eventually failed (failure to cross lesion) and in the remaining 2, a Monorail system solved the problem, which is incompatible with 4F catheters. In these 4 cases, a balloon could have been saved if the procedure had been started with a larger catheter and a movable wire system. We conclude that angioplasty through diagnostic 4F catheters completing a 4F coronary angiography is technically feasible and represents an economically viable alternative in selected patients. © 1993 Wiiey‐Liss, Inc.