A modification to the extraction-atomic-absorption method for the determination of antimony, bismuth, lead and tin
- 1 January 1974
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 99 (1181) , 469-470
- https://doi.org/10.1039/an9749900469
Abstract
It has been shown that the use of organometallic compounds for the calibration procedure used in the extraction-atomic-absorption determination of antimony, bismuth, lead and tin in metallurgical materials leads to erroneous results at low concentration levels. Results are presented, which demonstrate that the technique is satisfactory when calibrated with solutions that have been taken through the extraction procedure.Keywords