Large-area 200-ps gated microchannel plate detector

Abstract
Results are presented with a 15-mm-wide gated microchannel plate UV and x-ray detector. The active area is part of a 6-Ω transmission line driven by an electronically generated gate pulse. The microchannel plate is coated with CsI allowing tests with a frequency-quadrupoled, high-repetition-rate 1.05-μm laser. Results showing optical gate widths as short as 100 ps are presented.

This publication has 4 references indexed in Scilit: