Blood pressure measurement in children: the importance of cuff bladder size
- 1 October 1989
- journal article
- research article
- Published by Wolters Kluwer Health in Journal Of Hypertension
- Vol. 7 (10) , 845-850
- https://doi.org/10.1097/00004872-198910000-00010
Abstract
The effect of cuff bladder size on blood pressure measurement has been investigated in 838 children aged 5-7 years, using a Dinamap oscillometric automated blood pressure recorder and a Hawksley random zero sphygmomanometer. With both instruments the smallest (infant) cuffs recorded higher pressures than the child cuffs (mean systolic differences: Dinamap 4.6 mmHg, Hawksley 6.6 mmHg), which in turn recorded higher pressures than the largest (adult) cuffs (mean systolic differences: Dinamap 5.5 mmHg, Hawksley 8.1 mmHg). These differences in measurement between cuffs are sufficiently large as to be of concern on both clinical and epidemiological studies. Since they are little affected by arm circumference, their occurrence cannot be prevented simply by following current guidelines for selection of cuff size. Methods of dealing with these problems are discussed.Keywords
This publication has 6 references indexed in Scilit:
- BLOOD PRESSURE IN BRITISH CHILDREN: ASSOCIATIONS WITH ADULT BLOOD PRESSURE AND CARDIOVASCULAR MORTALITYThe Lancet, 1988
- ACCURACY OF BLOOD-PRESSURE MEASUREMENT BY THE DINAMAP MONITOR IN INFANTS AND CHILDREN1987
- Noninvasive determination of systolic, diastolic and end-systolic blood pressure in neonates, infants and young children: Comparison with central aortic pressure measurementsThe American Journal of Cardiology, 1983
- The error in indirect blood pressure measurement with the incorrect size of cuffAmerican Heart Journal, 1978
- The effect of cuff width on systolic blood pressure measurements in neonatesThe Journal of Pediatrics, 1977
- ACCURACY OF CLINICAL DETERMINATIONS OF BLOOD PRESSURE IN CHILDRENAmerican Journal of Diseases of Children, 1939