Separation of DNA sequencing fragments using an automated capillary electrophoresis instrument
- 1 January 1996
- journal article
- Published by Wiley in Electrophoresis
- Vol. 17 (1) , 144-151
- https://doi.org/10.1002/elps.1150170124
Abstract
Rapid, high‐resolution separation of DNA sequencing fragments by capillary gel electrophoresis using an automated, commercially available instrument is presented. The effect of column lengths and electric field strength on the resolution of sequencing fragments as well as the sensitivity of laser‐induced rescence (LIF) detection was investigated. Using a short capillary of 20 length, which results in a U‐shape of the capillary in the capillary cartridge very high separation efficiency, up to 17 × 106 theoretical plates per m, is obtained. Analysis of the band broadening factors revealed that the resolution on the short column is predominantly determined by axial diffusion and to a minor extent by detection zone width. Presumably due to the coiling of longer capillaries in the capillary cartridge, increasing the capillary length does increase the separation efficiency as predicted for diffusion‐limited separation. The concentration limit to detection (signal‐to‐noise ratio = 2) is 0.2 × 10−1M of fluorescein‐labeled oligonucleotide primer under the separating conditions for DNA sequencing samples. Increasing the electric field strength from 100 to 175 V/cm improved resolution and at the same time approximately doubled the sequencing speed. Fragments up to 500 nucleotides in length are resolved in less than 50 min.Keywords
This publication has 36 references indexed in Scilit:
- Effects of Linear Polyacrylamide Concentrations and Applied Voltages on the Separation of Oligonucleotides and DNA Sequencing Fragments by Capillary ElectrophoresisAnalytical Chemistry, 1994
- Separation of long DNA fragments by capillary gel electrophoresis with laser-induced fluorescence detectionElectrophoresis, 1994
- Optimization of electric field strength for DNA sequencing in capillary gel electrophoresisAnalytical Chemistry, 1993
- Influence of coiling on performance in capillary electrophoresis using open tubular and polymer network columnsJournal of Microcolumn Separations, 1992
- Quantitative analysis of DNA-sequencing electrophoresisGenetic Analysis: Biomolecular Engineering, 1992
- High Speed Automated DNA Sequencing in Ultrathin Slab GelsNature Biotechnology, 1992
- Why can we not sequence thousands of DNA bases on a polyacrylamide gel?Electrophoresis, 1992
- Three DNA sequencing methods using capillary gel electrophoresis and laser-induced fluorescenceAnalytical Chemistry, 1991
- Analysis of limiting factors of DNA band separation by a DNA sequencer using fluorescence detectionElectrophoresis, 1991
- The biased reptation model of DNA gel electrophoresis: Mobility vs molecular size and gel concentrationBiopolymers, 1989