Computerized Dosimetry of60Co Ophthalmic Applicators

Abstract
Theoretical dose distributions of 60Co ophthalmic applicators were calculated with the aid of a computer and the results were compared with experimental measurements obtained by thermoluminescent dosimetry and film emulsions in phantoms. The Sievert line-source method partially corrected for oblique filtration; however, a more precise dose description should include a program which completely corrects for oblique absorption in the ophthalmic applicator.

This publication has 0 references indexed in Scilit: