Correlation of microdosimetric measurements with relative biological effectiveness from clinical experience for two neutron therapy beams

Abstract
Microdosimetric measurements were made for the neutron therapy beams at the University of Chicago and at the Cleveland Clinic with the same geometry and phantom material using the same tissue‐equivalent spherical proportional counter and standard techniques. The energy deposition spectra (dose distributions in lineal energy) are compared for these beams and for their scattered components (direct beam blocked). The model of dual radiation action (DRA) of Kellerer and Rossi is employed to interpret these data in terms of biological effectiveness over this limited range of radiation qualities. The site‐diameter parameter of the DRA theory is determined for the Cleveland beam by setting the biological effectiveness (relative to 60Co gamma radiation) equal to the relative biological effectiveness value deduced from radiobiology experiments and clinical experience. The resulting value of this site‐diameter parameter is then used to predict the biological effectiveness of the Chicago beam. The prediction agrees with the value deduced from radiobiology and clinical experience. The biological effectiveness of the scattered components of both beams is also estimated using the model.
Funding Information
  • National Cancer Institute (CA 19720)