Mean energy in electron beams
- 1 September 1981
- journal article
- Published by Wiley in Medical Physics
- Vol. 8 (5) , 682-687
- https://doi.org/10.1118/1.595028
Abstract
The mean energy of the energy spectrum is an essential parameter for the dosimetry of therapeutic electron beams. Frequently it is assumed that the mean energy of such beams remains constant across the beam and only its degradation with depth is considered. The present work analyzes the variation of the mean energy of primary electrons with depth and lateral position in an electron beam using the Monte Carlo method. Results are compared with relations commonly employed for determination of mean energy at a depth. For the variation of the mean electron energy with depth in broad beams, good agreement was found between Monte Carlo results and an analytic continuous slowing down expression, which takes the variation of radiation stopping power with depth into account. Due to the calculated lateral variation of the mean energy, the relative absorbed dose profile determined with an air ionization chamber in a clinical beam should differ by less than 1% from the measured ionization profile.Keywords