Stabilization of Canine Pelvic Osteotomies with AO/ASIF Plates and Screws
- 1 January 1988
- journal article
- research article
- Published by Georg Thieme Verlag KG in Veterinary and Comparative Orthopaedics and Traumatology
- Vol. 01 (01) , 52-57
- https://doi.org/10.1055/s-0038-1633163
Abstract
Stabilization of pelvic osteotomy with a 2.7 mm dynamic compression plate (DCP) for the treatment of hip dysplasia, was performed in 42 operations in 27 dogs. Screw loosening was observed in 11 osteotomies (26%), plate bending in six cases (14%) and plate breakage in two cases (4%). An in vitro experiment using the pelvic bones of young mixed breed dogs showed no significant difference in the holding power of AO/ASIF 2.7 mm cortical, 3.5 mm cortical and 3.5 mm fully threaded cancellous screws in the body of the ilium. The 2.7 mm DCP has an advantage over the 3.5 DCP in stabilizing pelvic osteotomy because it allows more screws to be used per unit length of plate, which possibly results in a more stable fixation.Keywords
This publication has 4 references indexed in Scilit:
- Force measurements in screw fixationJournal of Biomechanics, 1976
- The Holding Power of Orthopedic Screws In VivoClinical Orthopaedics and Related Research, 1975
- A study of some factors which affect the strength of screws and their insertion and holding power in boneJournal of Biomechanics, 1968
- ACTUAL HOLDING POWER OF VARIOUS SCREWS IN BONEAnnals of Surgery, 1941